SHI Yan, XIAO Xin-yue, WEI Feng, et al. Determination of Harmful Elements in Cervi Cornus Colla Based on Microwave Digestion and ICP-AES Technology[J]. Chinese journal of experimental traditional medical formulae, 2013, 19(16): 151-153.
DOI:
SHI Yan, XIAO Xin-yue, WEI Feng, et al. Determination of Harmful Elements in Cervi Cornus Colla Based on Microwave Digestion and ICP-AES Technology[J]. Chinese journal of experimental traditional medical formulae, 2013, 19(16): 151-153. DOI: 10.11653/syfj2013160151.
Determination of Harmful Elements in Cervi Cornus Colla Based on Microwave Digestion and ICP-AES Technology
目的: 建立测定鹿角胶中铅、镉、砷、汞、铜和铬元素含量的方法。 方法: 采用微波消解和ICP-AES法同时测定11批市售鹿角胶中铅、镉、砷、汞、铜和铬元素的含量。微波消解条件为0~3 min
1 600 W
110℃;3~6 min
1 600 W
150℃;6~16 min
1 600 W
185℃。ICP-AES仪器测定条件为入射功率1.4 kW
冷却气氩气
冷却气流量12 L·min-1
载气流量1 L·min-1
辅助气流量0.8 L·min-1
样品提升量1 mL·min-1。 结果: 铅、镉、砷、汞、铜和铬元素检测限分别为0.15
0.01
0.27
0.08
0.03
0.03 mg·kg-1
加样回收率90%~112%
精密度、重复性均符合检测要求。各样品中镉、砷、汞元素均未检出
除F生产的样品中未检测到铅元素外
均检测出铅、铜和铬元素
其中部分样品中铬元素含量较高。 结论: 建立的方法准确、快捷
适用于胶类药材中有害元素的测定。
Abstract
Objective: To establish a method for determining the content of Pb
Cd
As
Hg
Cu and Cr in Cervi Cornus Colla. Method: Microwave digestion and inductively coupled plasma-atomic wmission spectrometry (ICP-AES) were adopted to determine the content of Pb
Cd
As
Hg
Cu and Cr in 11 batches of Cervi Cornus Colla.Microwave digestion conditions were as followings:0-3 min
1 600 W
110℃;3-6 min
1 600 W
150℃;6-16 min
1 600 W
185℃.Determination conditions of ICP-AES were:incident power of 1.4 kW
argon as cooling gas with flow rate of 12 L·min-1
flow rate of carrier gas and auxiliary with 1
0.8 L·min-1
volume of sample enhance 1 mL·min-1. Result: Detection limit of Pb
Cd
As
Hg
Cu and Cr were 0.15
0.01
0.27
0.08
0.03
0.03 mg·kg-1
respectively.recoveries was between 90%-112%
precision and repeatability were in line with detection requirments.Cd
As and Hg were not detected in each sample
apart from sample from F was not detected Pb
Pb
Cu and Cr were detected in other samples
the content of Cr was high in some samples. Conclusion: This method was accurate
fast and suitable for determination of harmful elements in glue herbs.