XIE Jun-ping, FENG Ling-ling, KUANG Xia. Determination of Methotrexate Level in Plasma Using Higher Doses Methotrexate by HPLC[J]. Chinese journal of experimental traditional medical formulae, 2010, 16(11): 69-72.
XIE Jun-ping, FENG Ling-ling, KUANG Xia. Determination of Methotrexate Level in Plasma Using Higher Doses Methotrexate by HPLC[J]. Chinese journal of experimental traditional medical formulae, 2010, 16(11): 69-72. DOI: 10.13422/j.cnki.syfjx.2010.11.070.
Determination of Methotrexate Level in Plasma Using Higher Doses Methotrexate by HPLC
摘要
目的:建立HPLC测定血清中甲氨蝶呤浓度的方法。方法:采用Gold色谱柱(4.6 mm×150 mm
Objective:To establish a HPLC method for determination of methotrexate concentration in plasma using higher doses.Method:The separation was performed in a Gold column(4.6 mm × 150 mm
5 μm) with a mobile phase of methanol:phosphate buffer(19 ∶ 81
pH 7.2) and determinated at 306 nm.Result:The measurable lowest concentration was 0.002 μmol.L-1.The experiment showed that the recovery from plasma was 90.2%
94.4%
101.2%.The RSD of within day and between day was less than 9.0%.A good linearity was found in the range of 0.01 ~ 1.38 μmol.L-1(r = 0.999 7) and in the range of 1.38 ~ 103.5 μmol.L-1(r = 0.999 5).Conclusion:The method is simple
sensitive and accurate with wide linear dynamic range.